Why Is The Deicing Time Of The Laboratory Snowflake Ice Machine Too Long?
Leave a message
Why is the deicing time of the laboratory snowflake ice machine too long?
The working process of the laboratory snowflake ice machine, from the beginning of ice making to the end of deicing,
The normal laboratory snowflake ice machine only takes 15 minutes to make ice, and it will be deiced after making it, which will not exceed 5 minutes;
Therefore, if a snowflake ice machine takes more than 5 minutes to de-ice, it can be considered that it has been de-icing for too long and there is a fault. What is the reason?
1. Malfunction of deicing solenoid valve
If the deicing solenoid valve is faulty, it is difficult to start the deicing program.
The inspection method is to try the heat supply pipe when deicing. If the heat supply pipe is not hot, it may be that the solenoid valve is faulty, or there is a problem with the relay controlling the solenoid valve on the circuit board.
2. There is foreign matter attached to the surface of the evaporator
If the adhesion of foreign matter on the surface of the evaporator results in a long deicing time, it is only necessary to clean the surface of the evaporator, the shower and the water storage box.
3. Refrigerant leakage
It is necessary to carefully check the ice machine to see where there is a leak, and then repair it and add refrigerant after identification.
4. Heating valve failure
If the heating valve is faulty and cannot be opened, the heating valve needs to be replaced.
The deicing time of the snowflake ice machine in the laboratory is too long, either the deicing solenoid valve is faulty, or there are foreign objects on the surface of the evaporator or refrigerant leakage, or the heating valve is faulty. Check one by one and solve it according to the method described in the article!